Specifications for our T-format TCSPC system

Excitation Wavelength Range:

Fundamental 740-980nm
Doubled 370-490nm
Tripled 247-327nm

Repetition Rate:

10MHz-1MHz and slower

Time Resolution Range (1024 Channels):

 

Picoseconds/channel
Collection Window
Lower Limit
3.26ps/ch
3.33ns
Upper Limit
1,950ps/ch
2000ns

Instrument Response Function:

FWHM 35.6ps

Sample Temperature Range:

-10°C to 105°C